P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms

Abstract

Background

Calibration of carotid artery tonometry waveforms needs to account for pulse pressure augmentation [1]. Due to its location nearer to the heart, carotid pulse pressure (carPP) is about 10 mmHg lower than brachial (brachPP) [1]. In practice, perfect applanation is hard to achieve with tonometry and thus tracking-errors may occur, leading to an exaggerated fall in the diastolic limb of the tonometry waveform. Consequently, the mean-to-diastolic difference of the waveform is underestimated, leading to an overestimation of carPP after calibration. We assessed the relationship between brachPP and carPP in 100 subjects, aged 18–80 yrs.

Methods and Results

Brachial pressure (Omron) and carotid tonometry (Sphygmocor) measurements were performed by a single, highly skilled vascular technician, according to Artery Society methodological standards. CarPP was 5 ± 13 mmHg lower than brachPP. Bland-Altman analysis revealed a positive trend (r = 0.695, p < 0.0001) between the carPP–brachPP difference and their mean, with carPP being 20 mmHg lower than brachPP at a mean 40 mmHg which extended to carPP being 20 mmHg higher(!) than brachPP at a mean 80 mmHg. Accordingly, 30% of the study population showed unphysiological carPP, indicating an unrealistic i.e. reverse amplification. Even with a tolerance of +5 mmHg, still 20% of cases showed unrealistic values.

Conclusion

In practice, calibrated carotid tonometry waveforms may yield unphysiological overestimations of carotid pulse pressure. This finding implies (potential) estimation errors in regression model studies on central pressure and local stiffness, suggesting brachial pulse pressure would be a valid compromise.

References

  1. Kelly R, Fitchett D. Noninvasive determination of aortic input impedance and external left ventricular power output: a validation and repeatability study of a new technique. J Am Coll Cardiol 1992;20:952–63.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Koen Reesink.

Rights and permissions

This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Reesink, K., Marais, L. & Boutouyrie, P. P122 Unrealistic Pulse Pressures from Calibrated Tonometric Waveforms. Artery Res 25 (Suppl 1), S162 (2019). https://doi.org/10.2991/artres.k-191224.148

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.2991/artres.k-191224.148