P144 Augmentation Index is not a Reliable Measure of Wave Reflections in Patients with Severe Aortic Stenosis

Abstract

Background

Augmentation index (AIx) serves as an independent predictor of cardiovascular events and mortality. It is often clinically used as a measure of arterial stiffness and reflections, a practice that was recently challenged. In this work, we sought to investigate the association between AIx and wave reflections in patients with severe aortic stenosis.

Methods

The final cohort comprised of 18 subjects (86 ± 4 years, 28% males) with severe aortic stenosis who underwent a Transcatheter Aortic Valve Implantation (TAVI). Invasive aortic and left ventricular pressure waveforms were obtained simultaneously before and after the TAVI. Key pressure waveform features were extracted and wave separation analysis was performed by combining aortic pressure and flow (Doppler echocardiography) data before and after the valve replacement.

Results

After the TAVI, aortic SBP and maximal blood flow velocity increased (from 134 ± 22 to 151 ± 30 mmHg, p = 0.04 and from 298 ± 65 to 376 ± 81 mL/sec, p < 0.01 respectively), while the ejection duration decreased (from 0.42 ± 0.07 to 0.35 ± 0.06 sec, p < 0.01). A significant decrease in AIx and augmented pressure was noted (from 44 ± 16 to 17 ± 12% and from 34 ± 13 to 17 ± 14 mmHg respectively, p < 0.01 for all). Wave separation analysis however showed a comparable increase in both forward (from 61 ± 14 to 79 ± 22 mmHg, p < 0.01) and backward (from 31 ± 11 to 39 ± 12 mmHg, p < 0.05) wave amplitudes. Reflection magnitude and backward wave transit time remained unchanged (p = NS).

Conclusion

Decrease in AIx after TAVI is not associated with changes in wave reflection amplitude or timing. For patients with severe aortic stenosis, interpretation of AIx should be done with caution.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Stamatia Pagoulatou.

Rights and permissions

This is an open access article distributed under the CC BY-NC 4.0 license (http://creativecommons.org/licenses/by-nc/4.0/).

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pagoulatou, S., Stergiopulos, N., Bikia, V. et al. P144 Augmentation Index is not a Reliable Measure of Wave Reflections in Patients with Severe Aortic Stenosis. Artery Res 25 (Suppl 1), S180 (2019). https://doi.org/10.2991/artres.k.191224.164

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.2991/artres.k.191224.164