P168 Feasibility Study of Local Pulse Wave Velocity Estimation in the Carotid Artery with Multi-Beam Laser Doppler Vibrometer

Abstract

Background

An innovative device using Laser Doppler Vibrometry (LDV) has been designed [1] to measure the transit time of the pulse wave between two locations along the course of the carotid artery (CA) from skin surface vibrations for assessment of local pulse wave velocity (PWV) [2].Aim: Tests were conducted on in-vitro models to assess the feasibility of the LDV to estimate the local PWV; preliminary in-vivo measurements were also performed.

Methods

Two CA geometries embedded within a soft-tissue-mimicking hydrogel were considered: i) a straight latex tube and ii) a patient-specific CA silicone-rubber model including the bifurcation. Models were pressurised in a water-filled loop and pulsatile flow was generated with a pump and/or high frequency impulses induced externally. For all measurements, two sets of six beams were used to measure surface displacement perpendicular to the external surface. PWV was calculated from the distance between selected beams and the delay between corresponding signals, using the time of the maximum of first and second derivatives of pressure (P-PWV) and displacement (LDV-PWV) as fiducial points [3]. A windowed cross correlation method [4] was also used for the in-vivo data analysis.

Results

PWV values for the in-vitro models are summarized in Table 1, while preliminary in-vivo LDV-PWV results are shown in Fig 1.

Conclusions

Good agreement between P-PWV and LDV-PWV in the tubular model was found under impulse loading, while complex waveforms measured under pulsatile flow and in-vivo conditions lead to more disparate effects when using different analysis methods. Further signal analysis is warranted.

Table 1 P-PWV and LDV-PWV mean values under pulsatile flow and induced impulse conditions.

PWV mean values and standard deviation for a young (24 years) and old (72 years) subject, systolic foot and dicortic notch were used as reference points. Note the higher PWV in the older subject, as expected.

PWV mean values and standard deviation from windowed cross correlation analysis around the dicrotic notch region.

References

  1. Y. Li, J. Zhu, M. Duperron, P. O’Brien, R. Schuler, S. Aasmul, M. De Melis, M. Kersemans, R. Baets. Six-beam homodyne laser Doppler vibrometry based on silicon photonics technology. Optics Express 2018, 26(3), p. 3638–3645.

    Google Scholar 

  2. T. Pereira, C Correia, J. Cardoso. Novel Methods: for Pulse Wave Velocity Measurement. Journal of Medical and Biological Engineering 2015, 35(5), 555–565.

    Google Scholar 

  3. E. Hermeling, KD. Reesink, LM. Kornmann, RS.Reneman, AP Hoeks. The dicrotic notch as alternative time-reference point to measure local pulse wave velocity in the carotid artery by means of ultrasonography. J Hypertens 2009, 27(10):2028–35.

    Google Scholar 

  4. T.G. Papaioannou, O. Vardoulis, A. Protogerou, G. Konstantonis, P. P. Sfikakis, C. Stefanadis and N. Stergiopulos. In vivo evaluation of a novel ‘diastole-patching’ algorithm for the estimation of pulse transit time: advancing the precision in pulse wave velocity measurement. Physiol Meas 2015, 36(1): 149–161.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This is an open access article distributed under the CC BY-NC license https://doi.org/creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tommasin, D., Li, Y., Reeves, J. et al. P168 Feasibility Study of Local Pulse Wave Velocity Estimation in the Carotid Artery with Multi-Beam Laser Doppler Vibrometer. Artery Res 24, 129–130 (2018). https://doi.org/10.1016/j.artres.2018.10.221

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1016/j.artres.2018.10.221