P38 Differences in Form Factor Calculated from Oscillometric or Waveform Mean Arterial Pressure

Abstract

Background

Oscillometric mean arterial pressure (MAP) agrees closely with invasive MAP, [1] but most devices do not report MAP and it is usually estimated by a form factor (FF). However, blood pressure (BP) measurement errors will affect FF, its correlations with exposures, and introduce errors into MAP estimated from the BP waveform.

Methods

Brachial BP was measured using a Pulsecor device in 1,112 participants in the Southall and Brent Revisited study (68.8 ± 6.1 y; 78.2% male; 47.4% White-European; 38.3% South-Asian; 14.3% African-Caribbean). Form factors (FFosc and FFwave) were calculated as (MAP-diastolic BP)/(systolic BP-diastolic BP) by oscillometry (MAPosc) or from the BP waveform (MAPwave).

Results

FFosc and FFwave differed (0.28 (SD = 0.02) vs. 0.36 (SD = 0.04); p <0.001) and were negligibly correlated (r = 0.07). Neither FFosc nor FFwave were associated with ethnicity, prevalent cardiovascular disease or current smoking status, and neither showed significant correlations with age, total-or HDL-cholesterol, or physical activity. Both FFosc and FFwave were lower in men (difference (Δ) = −0.005(95% CI = −0.007, −0.002) vs −0.015(95% CI = −0.020, 0.009) respectively) and were negatively correlated with height (r = −0.14 both), but only FFwave correlated with body mass index (r = 0.02 vs r = 0.10) and heart rate (r = −0.06 vs r = 0.20). ΔMAPosc-MAPwave correlated with age (r = 0.10), height (r = 0.15) and heart rate (0.17) and was greater in women (0.9(95% CI = 0.5, 1.3) mmHg).

Conclusions

FFwave agrees poorly with FFosc probably due to measurement errors. This creates spurious associations between exposures and FF and causes systematic errors in estimated MAPwave. These errors have the potential to confound associations in epidemiological studies.

References

  1. Ramsey M. Med Biol Eng Comput 1979;17:11–8.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Park, C., Tillin, T., Chaturvedi, N. et al. P38 Differences in Form Factor Calculated from Oscillometric or Waveform Mean Arterial Pressure. Artery Res 24, 90 (2018). https://doi.org/10.1016/j.artres.2018.10.091

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1016/j.artres.2018.10.091