- Research Article
- Open access
- Published:
A numerical study on the application of the functionally graded bioabsorbable materials in the stent design
Artery Research volume 24, pages 1–8 (2018)
Abstract
Background
Angioplasty with stenting is one of the primary treatment for coronary artery disease, hence, performance of the stent is deemed important. Bioabsorbable stents are the new generation of stents. Bioabsorbable magnesium alloys are a promising solution for adverse effects of long-term usage of stents. While their corrosion mechanics and biocompatibility have extensively been studied, there are no studies on application of functionally graded materials (FGM) on reduction of dogboning in bioabsorbable stents.
Methods
The objective of this study was at reducing the dogboning via the application of the FGM. A combination of the finite element (FE) method and optimization algorithm were employed to analyse/optimize the mechanical behaviour of bioabsorbable stents. Proposed FGM, in this study, was a combination of AZ80 and WE43 magnesium alloys. Dogboning of the FGM stent was chosen as objective function to be minimized and heterogeneous index was chosen as control variable in the optimization algorithm.
Results
The results revealed that the optimum FGM stent with heterogeneous index of 1.4625 has lower dogboning (63%) compared to that of the uniform stents made of AZ89 or WE43 magnesium alloy.
Conclusions
Furthermore, the results suggested that the plastic material properties have higher impact on the mechanical behaviour of the stent in comparison to the elastic material properties.
References
Buller CE, Dzavik V, Carere RG, Mancini GB, Barbeau G, Lazzam C. Primary stenting versus balloon angioplasty in occluded coronary arteries. Circulation 1999;100(3):236–42.
Karimi A, Navidbakhsh M, Shojaei A, Faghihi S. Measurement of the uniaxial mechanical properties of healthy and atherosclerotic human coronary arteries. Mater Sci Eng C 2013;33(5): 2550–4.
Boland EL, Shine R, Kelly N, Sweeney CA, McHugh PE. A review of material degradation modelling for the analysis and design of bioabsorbable stents. Ann Biomed Eng 2016;44(2):341–56.
Li H, Wang X. Design optimization of balloon-expandable coronary stent. Struct Multidiscip Optim 2013;48(4):837–47.
Khosravi A, Bahreinizad H, Bani MS, Karimi A. A numerical study on the application of the functionally graded materials in the stent design. Mater Sci Eng C 2017;73:182–8.
Gu X-N, Zheng Y-F. A review on magnesium alloys as biodegradable materials. Front Mater Sci China 2010;4(2):111–5.
Karimi A, Navidbakhsh M, Razaghi R. A finite element study of balloon expandable stent for plaque and arterial wall vulnerability assessment. J Appl Phys 2014;116(4). 044701–10.
Karimi A, Navidbakhsh M, Razaghi R. Plaque and arterial vulnerability investigation in a three-layer atherosclerotic human coronary artery using computational fluid-structure interaction method. J Appl Phys 2014;116(6). 064701–10.
Karimi A, Navidbakhsh M, Razaghi R, Haghpanahi M. A computational fluid-structure interaction model for plaque vulnerability assessment in atherosclerotic human coronary arteries. J Appl Phys 2014;115(14):144702–10.
Karimi A, Navidbakhsh M, Faghihi S, Shojaei A, Hassani K. A finite element investigation on plaque vulnerability in realistic healthy and atherosclerotic human coronary arteries. Proc Inst Mech Eng H 2013;227(2):148–61.
Karimi A, Razaghi R, Shojaei A, Navidbakhsh M. An experimental-nonlinear finite element study of a balloon expandable stent inside a realistic stenotic human coronary artery to investigate plaque and arterial wall injury. Biomed Tech (Berl) 2015;60(6):593–602.
Karimi A, Navidbakhsh M, Yamada H, Razaghi R. A nonlinear finite element simulation of balloon expandable stent for assessment of plaque vulnerability inside a stenotic artery. Med Biol Eng Comput 2014;52(7):589–99.
Miyamoto Y, Kaysser W, Rabin B, Kawasaki A, Ford RG. Functionally graded materials: design, processing and applications. Springer Science & Business Media; 2013.
Bever M, Duwez P. Gradients in composite materials. Mater Sci Eng C 1972;10:1–8.
Hedia H, Mahmoud NA. Design optimization of functionally graded dental implant. Bio Med Mater Eng 2004;14(2): 133–43.
Watanabe Y, Iwasa Y, Sato H, Teramoto A, Abe K, Miura-Fujiwara E. Microstructures and mechanical properties of titanium/biodegradable-polymer FGM for bone tissue fabricated by spark plasma sintering method. J Mater Process Technol 2011;211(12):1919–26.
Shirazi HA, Ayatollahi M. Biomechanical analysis of functionally graded biomaterial disc in terms of motion and stress distribution in lumbar spine. Int J Eng Sci 2014;84:62–78.
Dotter CT, Judkins MP. Transluminal treatment of arteriosclerotic obstruction: description of a new technic and a preliminary report of its application. Radiology 1989;172(3):904–20.
Roguin A. Stent: the man and word behind the coronary metal prosthesis. Circulation: Cardiovasc Interv 2011;4(2):206–9.
Lally C, Dolan F, Prendergast P. Cardiovascular stent design and vessel stresses: a finite element analysis. J Biomech 2005; 38(8):1574–81.
De Beule M, Mortier P, Carlier SG, Verhegghe B, Van Impe R, Verdonck P. Realistic finite element-based stent design: the impact of balloon folding. J Biomech 2008;41(2):383–9.
De Beule M, Van Impe R, Verhegghe B, Segers P, Verdonck P. Finite element analysis and stent design: reduction of dogboning. Technol Health Care 2006;14(4,5):233–41.
Wang W-Q, Liang D-K, Yang D-Z, Qi M. Analysis of the transient expansion behavior and design optimization of coronary stents by finite element method. J Biomech 2006;39(1):21–32.
Li N, Zhang H, Ouyang H. Shape optimization of coronary artery stent based on a parametric model. Finite Elem Anal Des 2009; 45(6):468–75.
Wu W, Petrini L, Gastaldi D, Villa T, Vedani M, Lesma E. Finite element shape optimization for biodegradable magnesium alloy stents. Ann Biomed Eng 2010;38(9):2829–40.
Powell MJ. The BOBYQA algorithm for bound constrained optimization without derivatives. Cambridge NA Report NA2009/06. Cambridge: University of Cambridge; 2009.
Khosrvai A, Akbari A, Bahreinizad H, Salimi Bani M, Karimi A. Optimizing through computational modeling to reduce dog-boning of functionally graded coronary stent material. J Mater Sci Mater Med 2017;28(9):142–50.
Lim D, Cho SK, Park WP, Kristensson A, Ko JY, Al-Hassani ST. Suggestion of potential stent design parameters to reduce restenosis risk driven by foreshortening or dogboning due to non-uniform balloon-stent expansion. Ann Biomed Eng 2008; 36(7):1118–29.
Karimi A, Navidbakhsh M, Shojaei A. A combination of histological analyses and uniaxial tensile tests to determine the material coefficients of the healthy and atherosclerotic human coronary arteries. Tissue Cell 2015;47(2):152–8.
Karimi A, Navidbakhsh M. A comparative study on the uniaxial mechanical properties of the umbilical vein and umbilical artery using different stress–strain definitions. Australas Phys Eng Sci Med 2014;37(4):645–54.
Karimi A, Sera T, Kudo S, Navidbakhsh M. Experimental verification of the healthy and atherosclerotic coronary arteries incompressibility via digital image correlation. Artery Res 2016;16:1–7.
Razaghi R, Karimi A, Rahmani S, Navidbakhsh M. A computational fluid–structure interaction model of the blood flow in the healthy and varicose saphenous vein. Vascular 2016;24(3): 254–63.
Karimi A, Rahmati SM, Sera T, Kudo S, Navidbakhsh M. A combination of constitutive damage model and artificial neural networks to characterize the mechanical properties of the healthy and atherosclerotic human coronary arteries. Artif Organs 2017;41(9):103–17.
Karimi A, Rahmati SM, Sera T, Kudo S, Navidbakhsh M. A combination of experimental and numerical methods to investigate the role of strain rate on the mechanical properties and collagen fiber orientations of the healthy and atherosclerotic human coronary arteries. Bioengineered 2017;8(2): 154–70.
Alagheband M, Rahmani S, Alizadeh M, Karimi A, Navidbakhsh M. Hemodynamic investigation of intraluminal thrombus effect on the wall stress in a stented three-layered aortic aneurysm model under pulsatile flow. Artery Res 2015;10:11–9.
Karimi A, Shojaei A, Razaghi R. Viscoelastic mechanical measurement of the healthy and atherosclerotic human coronary arteries using DIC technique. Artery Res 2017;18:14–21.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This is an open access article distributed under the CC BY-NC license https://doi.org/creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Bahreinizad, H., Bani, M.S., Khosravi, A. et al. A numerical study on the application of the functionally graded bioabsorbable materials in the stent design. Artery Res 24, 1–8 (2018). https://doi.org/10.1016/j.artres.2018.09.001
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1016/j.artres.2018.09.001