P.008 Comparative Study of Augmentation Index Measured by Two Different Devices
Artery Research volume 1, page 54 (2007)
Abstract
Background
Recent studies emphasize the significance of vessel wall properties estimations of large arteries not only for research purposes, but even inclinical practice already. The newly launched OMRON device measures the radial augmentation index (AIx) automatically. The aim of this study was to compare intra-individually this device to the long-term established Sphygmocor device in patients with manifest coronary heart disease and in addition to do an intra-device comparison of the OMRON device.
Methods
The investigated sample was one hundred randomly selected patients with manifest coronary heart disease, a random sub-sample of the Czech EuroAspire III study series. The measurements of the radial augmentation index were done in virtually similar conditions, in 60% twice using the HEM9000AI (OMRON) device and once the Sphygmocor (ATCor Systems) device, the rest of the sample was measured once by both devices.
Results
The OMRON device measured in average a higher AIx than the Sphygmocor device and this intraindividual difference is highly statistically significant. The median intraindividual difference between both devices was 17.4%. Both estimations significantly corelate, however, the correlation coefficient is for this kind of comparison rather low (about 0.44). In the OMRON intradevice comparison the median intraindividual difference was 1.9%.
Conclusion
The OMRON and Sphygmocor device estimations were in coronary patients not biologically equivalent. In conclusion, the OMRON device showed a very low variability and is applicable for estimation of the radial augmentation index in clinical use. However, the result from one device does not seem to be transferable to the other.
Author information
Authors and Affiliations
Rights and permissions
This is an open access article distributed under the CC BY-NC license https://doi.org/creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Patraulea, M., Mayer, O., Dolejsova, M. et al. P.008 Comparative Study of Augmentation Index Measured by Two Different Devices. Artery Res 1, 54 (2007). https://doi.org/10.1016/j.artres.2007.07.065
Published:
Issue Date:
DOI: https://doi.org/10.1016/j.artres.2007.07.065