P.077 In Vivo Validation of a Non-Invasively Measured Local Pulse Wave Velocity: Improved Performance Using Aortic Valve Closure in Stead of Aortic Valve Opening

Abstract

Introduction

High resolution multiple M-line ultrasound can be used to obtain local pulse wave velocity, by tracking the propagation of characteristic time-points of the distension waveform over an arterial segment. Local PWV provides a direct estimate of the local arterial stiffness.

Method

The left and right CCA of 12 young subjects was measured with multiple M-line ultrasound, resulting in 14 diameter waveforms spaced over 16.4mm. The second derivative of the distension waveforms was calculated to identify aortic valve opening and closure (AVO and AVC). By performing linear regression on M-line positions versus time-points, the local PWV was estimated. Beats were accepted for further analysis if the regression coefficient exceeded 0.9. The DC was measured using brachial-PP and carotid-distension, giving a reference estimate of the PWV using the Bramwell-Hill equation: \(PW{V_{DC}} = {1 \mathord{\left/ {\vphantom {1 {\sqrt {\rho DC} }}} \right. \kern-\nulldelimiterspace} {\sqrt {\rho DC} }}\)

Results

PWVAVC (mean = 4.91m/s) showed a better intra-subject precision (0.46m/s) than PWVAVO (mean = 5.21m/s, precision = 1.66m/s). The range of PWVDC (2.8–7.9m/s) is comparable to PWVAVC (3.5–7.3m/s) but not to PWVAVO (2.4–13.2m/s). The correlation between PWVDC and PWVAVC was weak but significant (r = 0.69, p<0.001). No correlation between PWVDC and PWVAVO exists.

Discussion

In contrast to PWVAVO, PWVAVC has an acceptable precision (coefficient of variation less than 10%) and correlates with PWVDC. The poor correlation between PWVAVO and PWVDC estimates may be caused by interference of early arterial wave-reflections leading to a biased estimate of PWVAVO. The correlation between PWVDC and PWVAVC is weak, although carotid-PP may improve DC calculation.

Author information

Authors and Affiliations

Authors

Rights and permissions

This is an open access article distributed under the CC BY-NC license https://doi.org/creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Cite this article

Hermeling, E., Reesink, K.D. & Hoeks, A.P. P.077 In Vivo Validation of a Non-Invasively Measured Local Pulse Wave Velocity: Improved Performance Using Aortic Valve Closure in Stead of Aortic Valve Opening. Artery Res 1, 70 (2007). https://doi.org/10.1016/j.artres.2007.07.011

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1016/j.artres.2007.07.011