From: Feasibility of Calculating Aortic Pulse Wave Velocity from Oscillometric Upper Arm Pulse Waves Using the Antares Algorithm

Fig. 2

Relationship between pulse wave velocity (PWV) values (n = 219) acquired by Antares algorithm (Antares PWV) and estimated pulse wave velocity (ePWV), based on mean blood pressure (MBP, Boutouyrie2). A Scatter plot for Antares PWV and ePWV (Boutouyrie2). Gray dotted line: linear regression line. Black dashed line: identity line. R2, coefficient of determination. B Bland–Altman plot for Antares PWV and ePWV (Boutouyrie2) with the representation of mean difference (black dashed line) and limits of agreement (black dotted line), from ± 1.96 SD. Mean difference ± SD: − 1.35 ± 1.24 m/s. Gray dotted line: linear regression line